Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 83: (1 year) CAL VER /5500 DATE: 27-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 65 NUMBER OF LINES: 233 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 80 Series Service Manual. # (PN 834168 April 1989, Rev. 3, 7/93, Change/Errata 8/94). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 83_85dsp.bmp # 55_8x_2W.bmp # 55_8x_CW.bmp # 55_8x_LA.bmp # 55_8x_HA.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 RSLT =DISPLAY TEST 1.006 PICE 83_85dsp 2.001 DISP No warm-up time requirement. 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 90%. 2.002 ASK- U 2.003 HEAD {ROTARY SWITCH TESTS} 2.004 DISP Hold down the RANGE button while turning the UUT rotary 2.004 DISP function switch from OFF to VDC. Release the RANGE 2.004 DISP button and turn the switch to {VAC}. 2.005 MATH MEM1 = 0 2.006 MEMI Enter UUT reading: 2.007 MEME 2.008 MEMC display 12U 3.001 DISP Rotate the UUT function switch to {VDC}. 3.002 MATH MEM1 = -32 3.003 MEMI Enter UUT reading: 3.004 MEME 3.005 MEMC display 12U 4.001 DISP Rotate the UUT function switch to {mVDC}. 4.002 MATH MEM1 = -64 4.003 MEMI Enter UUT reading: 4.004 MEME 4.005 MEMC display 12U 5.001 DISP Rotate the UUT function switch to {Ohms}. 5.002 MATH MEM1 = -96 5.003 MEMI Enter UUT reading: 5.004 MEME 5.005 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {Diode}. 6.002 MATH MEM1 = -128 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {mA/A}. 7.002 MATH MEM1 = -160 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {uA}. 8.002 MATH MEM1 = -192 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 ASK+ U 9.002 HEAD {AC VOLTAGE TESTS} 9.003 DISP Rotate the UUT function switch to OFF then VAC. 9.004 PIC 55_8x_2W 9.005 HEAD AC VOLTAGE TESTS: {400mV Range} 9.006 5500 400 350.0mV 3.9U 60H SI 2W 10.001 5500 400 350.0mV 5.7U 1kH SI 2W 11.001 5500 400 350.0mV 7.4U 5kH SI 2W 12.001 HEAD AC VOLTAGE TESTS: {4V Range} 12.002 5500 4 3.500V 0.038U 60H SI 2W 13.001 5500 4 3.500V 0.056U 1kH SI 2W 14.001 5500 4 3.500V 0.073U 5kH SI 2W 15.001 HEAD AC VOLTAGE TESTS: {40V Range} 15.002 5500 40 35.00V 0.38U 60H SI 2W 16.001 5500 40 35.00V 0.56U 1kH SI 2W 17.001 5500 40 35.00V 0.73U 5kH SI 2W 18.001 HEAD AC VOLTAGE TESTS: {400V Range} 18.002 5500 400 350.0V 3.8U 60H SI 2W 19.001 5500 400 350.0V 5.6U 1kH SI 2W 20.001 5500 400 350.0V 7.3U 5kH SI 2W 21.001 HEAD AC VOLTAGE TESTS: {1000V Range} 21.002 5500 1000 900V 12U 60H SI 2W 22.001 5500 1000 900V 26U 1kH SI 2W 23.001 5500 1000 900V 26U 5kH SI 2W 24.001 HEAD {FREQUENCY TESTS} 24.002 DISP Select the 400mV Range on the UUT. 24.002 DISP Press the Hz button to select the Frequency Counter. 24.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 24.004 5500 20 19.000kH 0.002U 150mV SI 2W 25.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 25.002 5500 200 190.00kH 0.02U 150mV SI 2W 26.001 HEAD {FREQUENCY SENSITIVITY AND TRIGGER LEVEL} 26.002 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V AC Range} 26.003 DISP Press the RANGE button to lock the 4V Range. 26.004 5500 1000.0H 30.0U 300mV SI 2W 27.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V DC Range} 27.002 DISP Rotate the UUT function switch to VDC. 27.002 DISP Press the RANGE button to lock the 4V Range. 27.002 DISP Press the Hz button to choose the Frequency mode. 27.003 5500 1000.0H 0.2U 1.7V SI 2W 28.001 5500 1000.0H 1.0V SI S 2W 28.002 EVAL Is the UUT {displaying all zeros}? 29.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {40V DC Range} 29.002 DISP Press the RANGE button to lock the 40V Range. 29.003 5500 1000.0H 0.2U 6.0V SI 2W 30.001 5500 1000.0H 2.0V SI S 2W 30.002 EVAL Is the UUT {displaying all zeros}? 31.001 HEAD {DC VOLTAGE TESTS} 31.002 DISP Press the Hz button twice to select VDC. 31.002 DISP 31.002 DISP Press the RANGE for 2s to select autorange. 31.003 HEAD DC VOLTAGE TESTS: {4V Range} 31.004 5500 4 3.500V 0.012U 2W 32.001 HEAD DC VOLTAGE TESTS: {40V Range} 32.002 5500 40 35.00V 0.12U 2W 33.001 5500 40 -35.00V 0.12U 2W 34.001 HEAD DC VOLTAGE TESTS: {400V Range} 34.002 5500 400 350.0V 1.2U 2W 35.001 HEAD DC VOLTAGE TESTS: {1000V Range} 35.002 5500 1000 1000V 4U 2W 36.001 HEAD {DC MILLIVOLT TEST} 36.002 DISP Rotate the UUT function switch to mVDC. 36.003 HEAD DC MILLIVOLT TEST: {400mV Range} 36.004 5500 400 350.0mV 1.2U 2W 37.001 HEAD {RESISTANCE TESTS} 37.002 DISP Rotate the UUT function switch to Ohms. 37.003 PIC 55_8x_CW 37.004 HEAD RESISTANCE TESTS: {400 Ohm Range} 37.005 5500 400 190.0Z 0.9U CW 38.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 38.002 5500 40 19.00kZ 0.09U CW 39.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 39.002 PIC 55_8x_2W 39.003 5500 4 1.900MZ 0.009U 2W 40.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 40.002 5500 40 19.00MZ 0.22U 2W 41.001 HEAD {CONDUCTANCE TEST} 41.002 DISP Press the Range button twice for 40nS range. 41.003 HEAD CONDUCTANCE TEST: {40nS Range} 41.004 5500 40 10.00nY 0.20U 2W 42.001 HEAD {CAPACITANCE TESTS} 42.002 DISP Press the BLUE button to select Capacitance. 42.003 HEAD CAPACITANCE TESTS: {5.00uF Range} 42.004 5500 5 1.00uF 0.03U 2W 43.001 HEAD CAPACITANCE TESTS: {0.500uF Range} 43.002 5500 0.5 0.470uF 0.007U 2W 44.001 HEAD CAPACITANCE TESTS: {0.0500uF Range} 44.002 5500 0.05 0.0470uF 0.0007U 2W 45.001 HEAD CAPACITANCE TESTS: {5.00nF Range} 45.002 DISP Select Relative mode (REL) to zero the display and 45.002 DISP automatically subtract the residual meter and test lead 45.002 DISP capacitance. 45.003 5500 5 4.70nF 0.07U 2W 46.001 HEAD {DIODE TEST} 46.002 DISP Rotate the UUT function switch to Diode. 46.003 5500 3.000V 0.061U 2W 47.001 DISP Disconnect the 5500A from the UUT. 47.002 HEAD {DC MILLIAMP TESTS} 47.003 DISP Rotate the UUT function switch to mA/A. 47.004 PIC 55_8x_LA 47.005 HEAD DC MILLIAMP TESTS: {40mA Range} 47.006 5500 40 35.00mA 0.16U 2W 48.001 HEAD DC MILLIAMP TESTS: {400mA Range} 48.002 5500 400 350.0mA 1.6U 2W 49.001 HEAD {AC MILLIAMP TESTS} 49.002 DISP Press the BLUE button to select AC. 49.003 HEAD AC MILLIAMP TESTS: {40mA Range} 49.004 5500 40 35.00mA 0.44U 60H SI 2W 50.001 5500 40 35.00mA 0.44U 1kH SI 2W 51.001 HEAD AC MILLIAMP TESTS: {400mA Range} 51.002 5500 400 350.0mA 4.4U 60H SI 2W 52.001 5500 400 350.0mA 4.4U 1kH SI 2W 53.001 HEAD {DC MICROAMP TESTS} 53.002 DISP Rotate the UUT function switch to uA. 53.003 HEAD DC MICROAMP TESTS: {400uA Range} 53.004 5500 400 350.0uA 1.7U 2W 54.001 HEAD DC MICROAMP TESTS: {4000uA Range} 54.002 5500 4000 3500uA 16U 2W 55.001 HEAD {AC MICROAMP TESTS} 55.002 DISP Press the BLUE button to select AC. 55.003 HEAD AC MICROAMP TESTS: {400uA Range} 55.004 5500 400 350.0uA 4.4U 60H SI 2W 56.001 5500 400 350.0uA 4.4U 1kH SI 2W 57.001 HEAD AC MICROAMP TESTS: {4000uA Range} 57.002 5500 4000 3500uA 44U 60H SI 2W 58.001 5500 4000 3500uA 44U 1kH SI 2W 59.001 HEAD {DC AMP TESTS} 59.002 DISP Rotate the UUT function switch to mA/A. 59.003 PIC 55_8x_HA 59.004 HEAD DC AMP TESTS: {4000mA Range} 59.005 5500 4000 3500mA 16U 2W 60.001 HEAD DC AMP TESTS: {10A Range} 60.002 5500 10 10.00A 0.06U 2W 61.001 HEAD {AC AMP TESTS} 61.002 DISP Press the BLUE button to select AC. 61.003 HEAD AC AMP TESTS: {4000mA Range} 61.004 5500 4000 3500mA 44U 60H SI 2W 62.001 5500 4000 3500mA 44U 1kH SI 2W 63.001 HEAD AC AMP TESTS: {10A Range} 63.002 5500 10 10.00A 0.14U 60H SI 2W 64.001 5500 10 10.00A 0.14U 1kH SI 2W 65.001 HEAD 65.002 END